#2721739
An cóimheas idir caighdeán seachtrach agus cainéal (CCSC) a chinneadh.
Determination of the external standard/channel ratio (ESCR).
An cóimheas idir caighdeán seachtrach agus cainéal (CCSC) a chinneadh.
Determination of the external standard/channel ratio (ESCR).
Z = an éifeachtúlacht a chomhfhreagraíonn don luach CCSC.
Z = the efficiency corresponding to the ESCR value.
Na samplaí ionsúiteachta ísle mar a thugtar tuairisc orthu i bpointe 6.4 ag a bhfuil luach CCSC 1,8 nó faoin tuairim sin, féadfar iad a thomhas tríd an CCSC, lena soláthraítear tomhas ar an gcóimheas éifeachtúlachta.
The low absorbance samples as described in point 6.4 with an ESCR value of approximately 1,8 may be measured through the ESCR, which provides a measure of the efficiency ratio.
Le linn an phróisis ina socraítear na cainéil (pointe 6.1), cinntear an CCSC trí úsáid a bhaint as an ríomhchlár iomchuí nuair a chinntear an éifeachtúlacht.
During the process of setting the channels (point 6.1) the ESCR is determined using the appropriate computer program when the efficiency is determined.
I gcás samplaí atá beagán múchta (CCSC thart ar 1,8), ní dhéantar ach difear diomaibhseach don éifeachtúlacht leis an athrú ar an luach seo.
In the case of slightly extinguished samples (ESCR circa 1,8) the efficiency is only negligibly affected by the change in this value.